Pressure testing machine,Universal testing machine,Constant stress pressure testing machine
· Liquid crystal display (LCD) universal testing machine series
· Liquid crystal display four pillar series universal testing machine
· The screen microcomputer universal testing machine series
· Display four pillar series microcomputer universal testing machine
· Microcomputer electro-hydraulic servo universal testing machine series
· Four pillar series microcomputer electro-hydraulic servo universal testing machine
A, An overview of the
Is mainly used for refractory material, cement block and masonry building materials such as the determination of the compressive strength, the use of hydraulic transmission, the electronic load and digital display, manual control load, the products comply with GB/T50081-2002 standard requirement.
This machine is advanced technology, value, high sensitivity, good stability, the new structure. Its reliable quality, convenient operation and good service. Is the building materials, construction, highway, railway, large and medium institutions and other units and departments of the necessary test equipment.
B, Installation, commissioning and maintenance
1.Testing machine should be installed in a clean, dry, temperature uniformity, no vibration, no corrosion The influence of corrosion gas in the laboratory.
2.Installation site location should be appropriate, should be easy to operation and maintenance, installation of foundation should be
Solid and reliable, the machine away from the wall should be around 800-1000 mm, so that the operation and maintenance.
3.Testing machine level adjustment, with 0.10/1000 mm level measurement oil cylinder
Surface, vertical and horizontal level + 1 case (or oblique pad iron pad under the base adjustment), good correction.
4.Oil oil is not used, due to the different climate will now oil specification is recommended
As follows: when the environment temperature of 15 ℃ + 5 ℃ recommended n46 GB433-84 hydraulic oil; When the environment temperature 28 ℃ + / - 5 ℃ n68 GB433-84 hydraulic oil as recommended.
5.When refueling after opening the cover plate, unscrewed the fuel cap a amount to the oil tank labeled (about 7 to 8 kg). Use after a period of time as found oil turbidity should be replaced. When changing unscrewed the tank bottom plug can drain the oil and add new oil into kerosene fuel tank for cleaning, and then released. So after a few repetitions, oil with a towel, at the bottom of adding new oil.
6.The machine should be kept clean inside and outside, to unprotected surface should be often besmear antirust oil, do not make
It should be to cut off the power supply with hood to cover up.
7.Installing a power supply, power supply should be paid attention to, whether the voltage is normal, not three-phase ac voltage
Lack of phase. Chassis grounding logo must be ground wire, open the power supply before, shall confirm the correct wiring.
8.Printer paper, the instrument will not be able to control print, please click the printer manual for Printing machine paper again.
This machine is high precision instruments and equipment, should be also responsible for, maintenance.
9.And according to certain cycle test precision. If found to have abnormal situation should be shutdown check immediately. If it is a major problem should please repair professional and technical personnel.
C, The main technical parameters
SKZY-300B |
Pressure testing machine |
Digital display level |
All rights reserved: wuxi city construction test instrument equipment co., LTD
Telephone: 400-860-8085 400-860-8085 Fax: 0510-8507-2875
Address: south long area in wuxi city, jiangsu province famous high-tech industrial park area C 31-1
The ICP registration number: zhejiang ICP for 05002238
Tech support: the league of network resources